COPE National  
    
5: STERILIZATION


In the past few years it has become apparent that the only acceptable standard for the instruments that you use in your practice is sterile. Why should you sterilize? Because you are using forceps and needles that enter the skin. Anything that goes below the skin must be sterile. Even though the heat used to remove hair is extremely hot, it does not heat long enough to sterilize.

There are several acceptable methods of sterilization that can be used. There is steam under pressure. There is dry heat. And there is chemical sterilization, either liquid or gas. For the next several issues of this news magazine, I will review the use of each of these methods of sterilization. The first, "Steam Under Pressure", is reviewed below.

STEAM UNDER PRESSURE

(THE AUTOCLAVE)

Steam under pressure is the most effective and widely used form of heat sterilization used today. It is both simple and inexpensive. The amount of heat that the air can carry is directly related to the amount of moisture present. The more moisture in the air, the more heat that can be carried. Steam under pressure is an effective carrier of heat. For example, a pressure cooker, cooks more quickly and effectively than an oven. There is one disadvantage, it cannot be used to sterilize heat and moisture sensitive items, powders or petroleum products. Steam kills microorganisms by first softening their protective coat or layer, which then will allow the heat to quickly coagulate the sensitive protein material inside. Steam under pressure requires four conditions in order to be effective. They are, contact, temperature, exposure time and moisture.

Any device being sterilized must be completely surrounded by steam. Steam being surface sterilant, needs direct contact with the microorganisms in order to be effective. To adequately sterilize, the temperature of the steam inside the sterilization chamber, must be at least 121 degrees Centigrade, or 250 degrees Fahrenheit throughout the entire exposure time. Devices must be exposed to the required temperature for a minimum period of time in order to kill microorganisms. The higher the temperature, the less time it takes. For sterilization to take place, there must be adequate moisture present in the sterilization chamber. The steam must be saturated, or in other words, reach a relative humidity of 100%.

Sterilization failures do occur, usually because all of the above four conditions have not been met. However, sterilization is also extremely difficult to achieve if the items being sterilized are improperly prepared, even when all of the conditions have been met. There are some frequent causes of sterilization failure. Inadequate cleaning of items, causing foreign matter to protect the microorganisms. Wrapping an item too tightly, hindering air removal and steam penetration of packages - air and steam do not mix; steam must replace the air in order for sterilization to take place. Overloading an autoclave will prevent the steam from surrounding all surfaces. Items may be positioned incorrectly so that the containers may trap air or water, for instance, pans, bottles or other airtight containers must have their tops removed and the containers placed on their sides.

In order to be assured that sterilization has taken place, the process must be monitored routinely. They are monitored biologically, mechanically, and chemically. The biological indicator for steam, is "Bacillus Stear-othermophilus spores", which are non pathogenic, but whose resistance and numbers are far greater than any microorganisms you would ever encounter. They present a greater sterilization challenge then your sterilizer would ever be faced with. Mechanical indicators include clocks, thermometers, pressure gauges and sometimes, recording devices. As most table top autoclave models do not have recording graphs or strips, it is recommended that you record the temperature or pressure indicated on the gauge yourself, at precise intervals during the sterilization cycle. If there are recording graphs or strips, they must be examined to ensure that the correct time, temperature and pressure has actually been reached during the cycle. Chemical indicators include chemically impregnated tapes and strips, or glass ampules containing indicator pellets. These can be most useful for table top sterilizers that rarely have built in recording devices. These indicators should be used for every sterilization load. A multi parameter process chemical indicator is probably the best type for these table top models. The indicators may be either internal or external. Internal indicators demonstrate by colour change that the inside of the package has been penetrated by steam. External indicators, like indicator tape, or markings on packaging materials, show by colour change that the item has been through a sterilization process, but not that it is actually sterile. Your equipment, procedures and techniques, will determine which chemical indicator will best suit your needs.

It doesn't matter what type of steam sterilizer you have. A traditional table top model, a table top with a cassette as the chamber, or one with a basket. The four conditions must be met for sterilization to take place, and the process must be monitored to assure a sterile product.

REFERENCES -

Association for the Advancement of Medical Instrumentation - "Good Hospital Practice: Steam Sterilization and Sterility Assurance" - 1988.

Canadian Standards Association - "Effective Sterilization in Hospitals by the Steam Process" - CAN/CSA - Z314.3-91, Toronto, Ontario, 1991.

Dornette, W.H.L.ed, - "Central Service Technical Manual", International Association of Hospital Central Service Management, 1981, Chicago, Illinois, pages 50 to 67.

Fallis, P.W. - "Handbook on Infection Control in Office-Based Health Care and Allied Services." - Canadian Standards Association, 1994, Toronto, Ontario - pages 67 to 75.

INFECTION CONTROL ARTICLES

  1. PREPARING FOR THE PATIENT/CLIENT
    Spring 2001, Volume 8, Number 1

  1. PREVENTING INFECTION FOR ELECTROLYSIS
    Fall 2000, Volume 7, Number 2

  1. INFLUENZA
    Spring 2000, Volume 7, Number 1

  1. MICROORGANISMS AND INFECTIOUS DISEASE
    Summer 1999, Volume 6, Number 2

  1. BLOOD BORNE INFECTIOUS DISEASES AND PERSONAL PROTECTION
    Spring 1999, Volume 6, Number 1

  1. MICROORGANISMS & INFECTION
    Fall 1998, Volume 5, Number 2.

  1. BIOLOGICAL TESTING OF YOUR STERILIZATION PROCESS
    Spring 1998, Volume 5, Number 1

  1. STEAM STERILIZATION
    Fall 1997, Volume 4, Number 2

  1. CHEMICAL STERILIZATION
    Spring 1997, Volume 4, Number1

  1. DRY HEAT STERILIZATION
    Fall 1996, Volume 3, Number 2

  1. STERILIZATION
    Spring 1996, Volume 3, Number 1

  1. IS YOUR USE OF NEEDLES SAFE IN YOUR PRACTICE?
    Fall 1995, Volume 2, Number 2

  1. PREPARING INSTRUMENTS FOR STERILIZATION
    Spring 1995, Volume 2, Number 1

  1. WHY ARE GLASS BEAD STERILIZERS NOT RECOMMENDED FOR USE IN ELECTROLOGY?
    Fall 1994, Volume 1, Number 2

  1. STERILIZATION: HOW, WHEN, AND WHAT WITH
    Spring 1994, Volume 1, Number1

 

Quick answers
 Find an electrologist
 Find an laser technician
 How to choose
 FAQ

Public Information
 History of Electrolysis
 Frequent Questions
 Who requires Electrolysis?
 Causes of Hair Growth
 Normal Hair Growth Cycle
 Electrolysis Procedure
 Why a Consultation?
 Treatment Programs
 Do-it-Yourself?
 Temporary Methods
 Electric Tweezer
 Laser Hair Removal

Infection Control Standards
 Introduction
 Routine Practices
 Infection Control Articles
 Public Health Inspection Information

 
Referral Information
Selecting an Electrologist
COPE's Referral Program

COPE Members
Electrolysis Technicians
Laser Technicians

Electrolysis Information
 Electrolysis Associations
 Training Information
 Classified Ads - Electrologists
 Supplier Ads
 Other Links

Membership Information
 Message from the Board
 COPE National Directors
 Membership Benefits
    Benefits Covered
    Benefits Not Covered
 Membership Requirements
 Membership Code of Ethics
 Membership Application Form
 Infection Control Contract

Continuing Educational
 Example Articles
 RDT Information

 RDT Application


Other Professional Services
 
Nacora Insurance

COPE Cancer Program

COPE Website
 Link to Us
 Contact Us
 Write the RDT Credential Exam
 Membership maintainence

Copyright © 1998-2007 COPE National.
Site maintainence by Jenny Lucas